Data record prod.megger-mwa330a.spec.turns-ratio-range
Verified 2026-10-27 Confidence: High
The manufacturer states the turns ratio accuracy of the MWA330A by ratio band at its highest excitation voltage.
Value: 80 V AC: ±0.1 % (0.8 to 2000), ±0.15 % (2001 to 4000), ±0.25 % (4001 to 10,000), ±0.30 % (10,001 to 45,000)
| Step | Item | Detail |
|---|---|---|
| Source | Megger datasheet | Manufacturer page |
| Data record | prod.megger-mwa330a.spec.turns-ratio-range | Verified 2026-10-27 |
| Entry | Megger MWA330A | No revisions |
- Subject
- Megger MWA330A (Products)
- Kind
- Value in context
- Source
- Megger datasheet, web page edition, MWA300/330A 3-Phase Ratio and Winding Resistance Analyzer data sheet (MWA3XX_DS_US_V06), Megger. Publisher page
History
| Date | Change |
|---|---|
| 2026-10-27 | created |
| 2026-10-27 | revised |
| 2026-10-27 | review_status unreviewed -> software_reviewed |
Cite this page
Cite the governing standard for the fact itself; cite this page as the finding aid.
"Data record prod.megger-mwa330a.spec.turns-ratio-range." Test Instrument Reference. 2026-10-27. https://www.testinstrumentreference.com/data/prod.megger-mwa330a.spec.turns-ratio-range/.
Test Instrument Reference, Data record prod.megger-mwa330a.spec.turns-ratio-range (Oct. 27, 2026), https://www.testinstrumentreference.com/data/prod.megger-mwa330a.spec.turns-ratio-range/.
@misc{etr-data-prod-megger-mwa330a-spec-turns-ratio-range,
title = {Data record prod.megger-mwa330a.spec.turns-ratio-range},
howpublished = {\url{https://www.testinstrumentreference.com/data/prod.megger-mwa330a.spec.turns-ratio-range/}},
organization = {Test Instrument Reference},
publisher = {Test Instrument Reference},
year = {2026},
note = {last substantive change 2026-10-27}
}{{cite web |title=Data record prod.megger-mwa330a.spec.turns-ratio-range |website=Test Instrument Reference |publisher=Test Instrument Reference |url=https://www.testinstrumentreference.com/data/prod.megger-mwa330a.spec.turns-ratio-range/ |date=2026-10-27 |access-date=}}