Data record instr.corona-camera.vs-infrared

Verified 2026-10-27 Confidence: High

Partial discharge is driven by voltage, not by heat, so ultraviolet detection does not depend on load and shows the stress even at low current, while an infrared camera needs enough load to produce detectable resistive heating; the ultraviolet image shows the stress before any heating and a thermal image then confirms whether resistive damage has followed.

ProvenanceOfil technology page supports data record instr.corona-camera.vs-infrared, which is cited on the entry Corona camera.SOURCEOfil technology pageOfil technology pageManufacturer pageDATA RECORDinstr.corona-camera.vs-infraredinstr.corona-camera.vs-i..Verified 2026-10-27ENTRYCorona cameraCorona cameraNo revisions
Provenance. Where this record comes from and where it is used: the cited source, the record itself and the entry that cites it.
Provenance: the chain from source to entry.
StepItemDetail
SourceOfil technology pageManufacturer page
Data recordinstr.corona-camera.vs-infraredVerified 2026-10-27
EntryCorona cameraNo revisions
Subject
Corona camera (Instrument classes)
Kind
Instrument fact
Source
Ofil technology page, web page edition, Partial Discharge and Corona Detection, Ofil. Publisher page

History

Every recorded change to this data record, oldest first.
DateChange
2026-10-27created
2026-10-27revised
2026-10-27review_status unreviewed -> software_reviewed

Cite this page

Cite the governing standard for the fact itself; cite this page as the finding aid.

"Data record instr.corona-camera.vs-infrared." Test Instrument Reference. 2026-10-27. https://www.testinstrumentreference.com/data/instr.corona-camera.vs-infrared/.