Data record instr.cable-fault-locator.fault-classification

Verified 2026-10-27 Confidence: High

The first measurement stage identifies the faulted phase and whether the fault is of low or high resistance, because this decides the technique: a low-voltage pulse from a time-domain reflectometer sees a low-resistance fault, while a high-resistance fault is not visible to it and needs a surge (impulse) generator.

ProvenanceMegger application page supports data record instr.cable-fault-locator.fault-classification, which is cited on the entry Cable fault locator.SOURCEMegger application pageMegger application pageManufacturer pageDATA RECORDinstr.cable-fault-locator.fault-classificationinstr.cable-fault-locato..Verified 2026-10-27ENTRYCable fault locatorCable fault locatorNo revisions
Provenance. Where this record comes from and where it is used: the cited source, the record itself and the entry that cites it.
Provenance: the chain from source to entry.
StepItemDetail
SourceMegger application pageManufacturer page
Data recordinstr.cable-fault-locator.fault-classificationVerified 2026-10-27
EntryCable fault locatorNo revisions
Subject
Cable fault locator (Instrument classes)
Kind
Instrument fact
Source
Megger application page, web page edition, Cable fault location, Megger. Publisher page

History

Every recorded change to this data record, oldest first.
DateChange
2026-10-27created
2026-10-27review_status unreviewed -> software_reviewed

Cite this page

Cite the governing standard for the fact itself; cite this page as the finding aid.

"Data record instr.cable-fault-locator.fault-classification." Test Instrument Reference. 2026-10-27. https://www.testinstrumentreference.com/data/instr.cable-fault-locator.fault-classification/.