Data record gloss.absorption-current.usage
Verified 2026-10-27 Confidence: High
In timed insulation tests the absorption current falls more slowly than the capacitive charging current, and on good insulation the displayed resistance keeps rising as the charging currents disappear. On poor insulation a high leakage current hides the charging currents, so a graph of resistance against time is much flatter.
| Step | Item | Detail |
|---|---|---|
| Source | Megger Electrical Tester article | Manufacturer page |
| Data record | gloss.absorption-current.usage | Verified 2026-10-27 |
| Entry | Absorption current | No revisions |
- Subject
- Absorption current (Glossary)
- Kind
- Definition
- Source
- Megger Electrical Tester article, 2008-10-16 edition, Insulation testing diagnostic techniques, Megger. Publisher page
History
| Date | Change |
|---|---|
| 2026-10-27 | created |
| 2026-10-27 | review_status unreviewed -> software_reviewed |
Cite this page
Cite the governing standard for the fact itself; cite this page as the finding aid.
"Data record gloss.absorption-current.usage." Test Instrument Reference. 2026-10-27. https://www.testinstrumentreference.com/data/gloss.absorption-current.usage/.
Test Instrument Reference, Data record gloss.absorption-current.usage (Oct. 27, 2026), https://www.testinstrumentreference.com/data/gloss.absorption-current.usage/.
@misc{etr-data-gloss-absorption-current-usage,
title = {Data record gloss.absorption-current.usage},
howpublished = {\url{https://www.testinstrumentreference.com/data/gloss.absorption-current.usage/}},
organization = {Test Instrument Reference},
publisher = {Test Instrument Reference},
year = {2026},
note = {last substantive change 2026-10-27}
}{{cite web |title=Data record gloss.absorption-current.usage |website=Test Instrument Reference |publisher=Test Instrument Reference |url=https://www.testinstrumentreference.com/data/gloss.absorption-current.usage/ |date=2026-10-27 |access-date=}}