Clamp-on ground tester

Updated 2026-10-27

Data records

  • Instruments of this class store readings and carry separate user-set alarm limits for resistance, voltage and current, so that a reading outside the expected band is flagged during a survey of many electrodes. Source: AEMC product page, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.alarms-and-logging
  • A clamp-on ground tester also reads the alternating current flowing in the grounding conductor, so leakage current is observed without disconnecting the electrode; a selectable band-pass filter restricts the reading to power-frequency current. Source: Fluke product page, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.leakage-current
  • The value displayed is the resistance of the loop enclosed by the clamp, not an isolated electrode resistance; where equipment bonding creates a metallic loop in the ground system, the instrument reads that loop and the stakeless technique cannot be used at that point. Source: Megger product page, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.loop-not-individual
  • Because the electrode stays connected and no stakes are driven, the clamp method is used where stakes cannot be placed, such as inside buildings, on paved sites and on transmission structures, and it avoids the hazard of breaking a ground connection on an operating system. Source: Fluke product page, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.no-disconnection
  • A clamp-on tester cannot measure an isolated electrode, because a single path to ground gives the induced current no closed circuit; the fall-of-potential method is used instead. Source: Fluke application article, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.not-applicable
  • A clamp-on ground tester induces a known voltage into the grounding conductor through one clamp winding and measures the current that results through a second winding, or the second half of a dual clamp, and computes the resistance of the loop that the conductor forms with the ground system. Source: Fluke application article, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.principle
  • The method requires the electrode under test to be one of several parallel grounds, because the induced current must return through the other ground paths; the reading treats the combined resistance of those parallel paths as negligible against the single path under test. Source: Fluke application article, web page edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.return-path
  • The instrument class is written Clamp-on ground tester. Value: Clamp-on ground tester. Source: Electrical Testing Reference editorial taxonomy, 2026-10 edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.title
  • Because the clamp method gives no independent check of its own result, maker guidance treats it as a trending tool for an existing ground system rather than as a verification of a new installation that has no earlier results for comparison. Source: Megger Electrical Tester article, 2007-11-01 edition. Verified 2026-10-27. Record instr.clamp-on-ground-tester.trending

Sources

  • AEMC product page, Ground Resistance Tester Model 6417, web page edition, AEMC Instruments. Held as: Public page. Publisher page
  • Electrical Testing Reference editorial taxonomy, Electrical Testing Reference editorial taxonomy of test methods, 2026-10 edition, Electrical Testing Reference. Held as: Own observation.
  • Fluke application article, Stakeless method of earth grounding, web page edition, Fluke. Held as: Public page. Publisher page
  • Fluke product page, Fluke 1630-2 FC Earth Ground Clamp, web page edition, Fluke. Held as: Public page. Publisher page
  • Megger Electrical Tester article, A guide to earth testing, 2007-11-01 edition, Megger. Held as: Public page. Publisher page
  • Megger product page, DET4 series four-terminal earth/ground resistance testers, web page edition, Megger. Held as: Public page. Publisher page
Changelog (18 entries)
DateRevisionChange
2026-10-27created (instr.clamp-on-ground-tester.alarms-and-logging)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.alarms-and-logging)
2026-10-27created (instr.clamp-on-ground-tester.leakage-current)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.leakage-current)
2026-10-27created (instr.clamp-on-ground-tester.loop-not-individual)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.loop-not-individual)
2026-10-27created (instr.clamp-on-ground-tester.no-disconnection)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.no-disconnection)
2026-10-27created (instr.clamp-on-ground-tester.not-applicable)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.not-applicable)
2026-10-27created (instr.clamp-on-ground-tester.principle)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.principle)
2026-10-27created (instr.clamp-on-ground-tester.return-path)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.return-path)
2026-10-27created (instr.clamp-on-ground-tester.title)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.title)
2026-10-27created (instr.clamp-on-ground-tester.trending)
2026-10-27review_status unreviewed -> software_reviewed (instr.clamp-on-ground-tester.trending)